The effects of lateral head tilt on ocular astigmatic axis

Authors

Isfahan Eye Research Center, Isfahan University of Medical Sciences, Isfahan, Iran

Abstract

Background: Compensatory ocular counter-torsion (COCT) is supposed to maintain the eyes aligned with the visual environment following head tilt. Because of some recent controversies the functional capacity of this phenomenon was defined according to the extent of induced astigmatic axis error following head tilt.
Materials and Methods: Objective autorefractometry was performed on 70 eyes with a regular astigmatism of ≥2D at vertical, right head tilt and left head tilt positions of 5°, 10°, 15°, 20° and 25°. Astigmatic axis error was calculated according to the difference between the defined axis at each tilted head position and the defined axis at the vertical head position. A tiltometer was used for this purpose to show the angle of head tilt without disturbing the process of refractometry.
Results: The mean astigmatic axis error was 3.2° ± 1.5° and 18.4° ± 4.2° at the head tilt angles of 5° and 25° respectively. The mean percentage of tilt angle compensation by COCT was 36% and 26% at the head tilt angles of 5° and 25° respectively. There was a direct relation between the head tilt angle and the induced astigmatic axis error (ANOVA, P < 0.001, 95% of confidence interval [CI]). Astigmatic axis error values at right head tilt were significantly lower than their corresponding values at left head tilt (ANOVA, P = 0.04 95% CI).
Conclusion: Any minimal angle of head tilt may cause erroneous measurement of astigmatic axis and should be avoided during refraction. One cannot rely on the compensatory function of ocular counter-torsion during the refraction.

Keywords

1. Jampel RS, Shi DX. The absence of so-called compensatory ocular countertorsion: The response of the eyes to head tilt. Arch Ophthalmol 2002;120:1331-40.  Back to cited text no. 1
[PUBMED]    
2. Simonsz HJ. The history of the scientific elucidation of ocular counterrolling. Doc Ophthalmol 1985;61:183-9.  Back to cited text no. 2
[PUBMED]    
3. Kushner BJ. Ocular torsion: Rotations around the "WHY" axis. J AAPOS 2004;8:1-12.  Back to cited text no. 3
[PUBMED]    
4. Fesharaki H. Tiltometer. Geneva inventions, 40 th international inventions of Geneva, World Intellectual Property Organization (WIPO); 2012; April 18-22. Official Catalogue: 127.   Back to cited text no. 4
    
5. Nagel A. On the occurrence of true rolling of the eye about the line of sight. 2. Strabismus 2000;8:33-8.  Back to cited text no. 5
[PUBMED]    
6. Adler FH. Physiologic factors in differential diagnosis of paralysis of superior rectus and superior oblique muscles. Arch Ophthal 1946;36:661-73.  Back to cited text no. 6
[PUBMED]    
7. Schworm HD, Ygge J, Pansell T, Lennerstrand G. Assessment of ocular counterroll during head tilt using binocular video oculography. Invest Ophthalmol Vis Sci 2002;43:662-7.  Back to cited text no. 7
[PUBMED]    
8. Pansell T, Tribukait A, Bolzani R, Schworm HD, Ygge J. Drift in ocular counterrolling during static head tilt. Ann N Y Acad Sci 2005;1039:554-7.  Back to cited text no. 8
[PUBMED]    
9. Sverkersten U, Ygge J, Pansell T. Drift of visually induced optostatic torsion. Strabismus 2009;17:82-9.  Back to cited text no. 9
[PUBMED]    
10. Palla A, Bockisch CJ, Bergamin O, Straumann D.Dissociated hysteresis of static ocular counterroll in humans. J Neurophysiol. 2006 Apr;95:2222-32.  Back to cited text no. 10