Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification

Authors

1 Department of Medical Physics, School of Medicine, Isfahan University of Medical Sciences, Isfahan; Department of Medical Physics, School of Medicine, Rafsanjan University of Medical Sciences, Rafsanjan; Department of Medical Physics, School of Medicine, Hamadan University of Medical Sciences, Hamadan, Iran

2 Department of Medical Physics, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran

3 Department of Radiation Oncology, Isfahan Milad Hospital, Isfahan, Iran

4 Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, The Netherlands

Abstract

Background: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans, both pretreatment and in vivo. The aim of this study is to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in vivo as well.
Materials and Methods: Dose distributions were reconstructed from EPID images using a 2D EPID dosimetry algorithm inside a homogenous slab phantom for a simple 10 × 10 cm2 box technique, 3D conformal (prostate, head-and-neck, and lung), and intensity-modulated radiation therapy (IMRT) prostate plans inside an anthropomorphic (Alderson) phantom and in the patients (one fraction in vivo) for 3D conformal plans (prostate, head-and-neck and lung).
Results: The planned and EPID dose difference at the isocenter, on an average, was 1.7% for pretreatment verification and less than 3% for all in vivo plans, except for head-and-neck, which was 3.6%. The mean γ values for a seven-field prostate IMRT plan delivered to the Alderson phantom varied from 0.28 to 0.65. For 3D conformal plans applied for the Alderson phantom, all γ1% values were within the tolerance level for all plans and in both anteroposterior and posteroanterior (AP-PA) beams.
Conclusion: The 2D EPID-based dosimetry algorithm provides an accurate method to verify the dose of a simple 10 × 10 cm2 field, in two dimensions, inside a homogenous slab phantom and an IMRT prostate plan, as well as in 3D conformal plans (prostate, head-and-neck, and lung plans) applied using an anthropomorphic phantom and in vivo. However, further investigation to improve the 2D EPID dosimetry algorithm for a head-and-neck case, is necessary.

Keywords

1.
Steinke MF, Bezak E. Technological approaches to in-room CBCT imaging. Australas Phys Eng Sci Med 2008;31:167-79.  Back to cited text no. 1
    
2.
vanHerk M. Different styles of image-guided radiotherapy. Semin Radiat Oncol 2007;17:258-67.  Back to cited text no. 2
    
3.
Dawson LA, Jaffray DA. Advances in image-guided radiation therapy. J Clin Oncol 2007;25:938-46.  Back to cited text no. 3
    
4.
Mans A, Wendling M, McDermott LN, Sonke JJ, Tielenburg R, Vijlbrief R, et al. Catching errors with in vivo EPID dosimetry. Med Phys 2010;37:2638-44.  Back to cited text no. 4
    
5.
Wendling M, Louwe RJ, McDermott LN, Sonke JJ, van Herk M, Mijnheer BJ. Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method. Med Phys 2006;33:259-73.  Back to cited text no. 5
    
6.
Kirby MC, Glendinning AG. Developments in electronic portal imaging systems. Br J Radiol 2006;79:S50-65.  Back to cited text no. 6
    
7.
Kaatee RS, Olofsen MJ, Verstraate MB, Quint S, Heijmen BJ. Detection of organ movement in cervix cancer patients using a fluoroscopic electronic portal imaging device and radiopaque markers. Int J Radiat Oncol Biol Phys 2002;54:576-83.  Back to cited text no. 7
    
8.
Mohammadi M, Bezak E. The physical characteristics of a SLIC-EPID for transmitted dosimetry. Int J Radiat Res 2005;2:175-83.  Back to cited text no. 8
    
9.
Nijsten SM, Mijnheer BJ, Dekker AL, Lambin P, Minken AW. Routine individualised patient dosimetry using electronic portal imaging devices. Radiother Oncol 2007;83:65-75.  Back to cited text no. 9
    
10.
Erridge SC, Seppenwoolde Y, Muller SH, van Herk M, De Jaeger K, Belderbos JS, et al. Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer. Radiother Oncol 2003;66:75-85.  Back to cited text no. 10
    
11.
Hurkmans CW, Remeijer P, Lebesque JV, Mijnheer BJ. Set-up verification using portal imaging; review of current clinical practice. Radiother Oncol 2001;58:105-20.  Back to cited text no. 11
    
12.
Van den Heuvel F, Fugazzi J, Seppi E, Forman JD. Clinical application of a repositioning scheme, using gold markers and electronic portal imaging. Radiother Oncol 2006;79:94-100.  Back to cited text no. 12
    
13.
Bogaerts R, Van Esch A, Reyman R, Huyskens D. A method to estimate the transit dose on the beam axis for verification of dose delivery with portal images. Radiother Oncol 2000;54:39-46.  Back to cited text no. 13
    
14.
McDermott LN, Wendling M, van Asselen B, Stroom J, Sonke JJ, van Herk M, et al. Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification. Med Phys 2006;33:3921-30.  Back to cited text no. 14
    
15.
McDermott LN, Wendling M, Sonke JJ, van Herk M, Mijnheer BJ. Replacing pretreatment verification with in vivo EPID dosimetry for prostate IMRT. Int J Radiat Oncol Biol Phys 2007;67:1568-77.  Back to cited text no. 15
    
16.
McDermott LN, Wendling M, Nijkamp J, Mans A, Sonke JJ, Mijnheer BJ, et al. 3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT. Radiother Oncol 2008;86:35-42.  Back to cited text no. 16
    
17.
Van Elmpt W, McDermott L, Nijsten S, Wendling M, Lambina P, Mijnheer B. A literature review of electronic portal imaging for radiotherapy dosimetry. Radiother Oncol 2008;88:289-309.  Back to cited text no. 17
    
18.
Pasma KL, Dirkx ML, Kroonwijk M, Visser AG, Heijmen BJ. Dosimetricverification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device. Med Phys 1999;26:2373-8.  Back to cited text no. 18
    
19.
McNutt TR, Mackie TR, Reckwerdt P, Papanikolaou N, Paliwal BR. Calculation of portal dose using the convolution/superposition method. Med Phys 1996;23:527-35.  Back to cited text no. 19
    
20.
Huyskens D, Van Dam J, Dutreix A. Midplane dose determination using in vivo dose measurements in combination with portal imaging. Phys Med Biol 1994;39:1089-101.  Back to cited text no. 20
    
21.
Terrón JA, Sánchez-Doblado F, Arráns R, Sánchez-Nieto B, Errazquin L. Midline dose algorithm for in vivo dosimetry. Med Dosim 1994;19:263-7.  Back to cited text no. 21
    
22.
Rizzotti A, CompriC, Garusi GF. Dose evaluation to patients irradiated by 60Co beams, by means of direct measurement on the incident and on the exit surfaces. Radiother Oncol 1985;3:279-83.  Back to cited text no. 22
    
23.
Boellaard R, Essers M, van Herk M, Mijnheer BJ. New method to obtain the midplane dose using portal in vivo dosimetry. Int J Radiat Oncol Biol Phys 1998;41:465-74.  Back to cited text no. 23
    
24.
Jordan TJ, Williams PC. The design and performance characteristics of a multileaf collimator. Phys Med Biol 1994;39:231-51.  Back to cited text no. 24
    
25.
McDermott LN, Louwe RJ, Sonke JJ, van Herk MB, Mijnheer BJ. Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device. Med Phys 2004;31:285-95.  Back to cited text no. 25
    
26.
Louwe RJ, McDermott LN, Sonke JJ, Tielenburg R, Wendling M, van Herk MB, et al. The long-term stability of amorphous silicon flat panel imaging devices for dosimetry purposes. Med Phys 2004;31:2989-95.  Back to cited text no. 26
    
27.
Boellaard R, van Herk M, Uiterwaal H, Mijnheer B. First clinical tests using a liquid-filled electronic portal imaging device and a convolution model for the verification of the midplane dose. Radiother Oncol 1998;47:303-12.  Back to cited text no. 27
    
28.
Louwe RJ, Damen EM, van Herk M, Minken AW, Törzsök O, Mijnheer BJ. Three-dimensional dose reconstruction of breast cancer treatment using portal imaging. Med Phys 2003;30:2376-89.  Back to cited text no. 28
    
29.
Wendling M, McDermott LN, Mans A, Olaciregui-Ruiz Í, Pecharromán-Gallego R, Sonke JJ, et al. In aqua vivo EPID dosimetry. Med Phys 2012;39:367-77.  Back to cited text no. 29
    
30.
Pecharromán-Gallego R, Mans A, Sonke JJ, Stroom JC, Olaciregui-Ruiz I, van Herk M, et al. Simplifying EPID dosimetry for IMRT treatment verification. Med Phys 2011;38:983-92.  Back to cited text no. 30
    
31.
vanElmpt WJ, Nijsten SM, Dekker AL, Mijnheer BJ, Lambin P. Treatment verification in the presence of inhomogeneities using EPID-based three-dimensional dose reconstruction. Med Phys 2007;34:2816-26.  Back to cited text no. 31